K15 Touchscreen

multi-parameter monitor - ECG, RESP, TEMP, NIBP, SpO2, IBP, EtCO2 optional

K15 Touchscreen

multi-parameter monitor - ECG, RESP, TEMP, NIBP, SpO2, IBP, EtCO2 optional

Code:
Brand:
Unit:
Availability:

2,331.00 €

(including VAT)

Price without VAT 2,220.00 €
info General information
assignment Technical information
work Standard equipment
list Compatible with
save_alt Downloads

General information

Multifunctional monitor to monitor the vital signs of adult, pediatric and neonatal patients. With the functions of real-time recording and display of parameters such as ECG, HR heart rate, NIBP non-invasive blood pressure, SpO2 saturation, RESP respiration, TEMP temperature, IBP invasive blood pressure.
  • Ergonomic design with intuitive user interface
  • USB socket for data transfer and software update
  • 9 traces on video waveforms
  • View in Windows environment, easy to manage
  • Visual and audible alarm for physiological and technical anomalies
  • Large memory capacity
  • Suitable for use with genuine Nellcor sensors
  • Protection against defibrillator discharges, resistance against electrosurgical unit interference; cardiac pacemaker pulse detection and suppression
Internal software: GB, DE, FR, PT, ES, IT, PL, TR, RU.
PC software GB, IT, ES. Compatible with Windows XP, 7, 8,10, 11.

Standard equipment

  • SpO2 adult probe
  • Adult NIBP cuff
  • SpO2 extension cable
  • ECG cable
  • Skin temperature probe
  • Power cord
  • Lithium rechargeable battery
  • Dust cover
  • Disposable electrode
  • Manual: GB, FR, IT, ES, PL, RU
  • IBP module 2, cable, sensor
Printer not included. If ordered together with the monitor, the installation is carried out by our technician; in case it is ordered later, the installation requires an external technician (to be paid by the customer).



Technical information

ECG
  • Input dynamic range: ±(0.5~5mVp)
  • Differential input impedance: =10 MO
  • Bandwidth:
    - 0.05~150Hz (Diagnostics)
    - 0.5~40Hz (Monitor)
    - 1~20Hz (Operation)
  • CMRR:
    - =90dB (Diagnostics)
    - =105 dB (Monitoring & Operation)
  • Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto
  • Scan Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
  • HR measuring range: 15~350bpm ±1% or ±2bpm, whichever is greater
  • Pacemaker pulse detection and inhibition function
RESP
  • Measuring range: 0~120rpm ±5% or ±2rpm, whichever is greater
TEMP
  • Measuring range: 21.0~50.0°C ±0.2°C between 25~45°C
NIBP
  • Technique: Oscillometric method
  • Typical measurement time: <30 seconds (adult cuff)
  • NIBP SYS measuring range:
    - 40~275 mmHg (Adult)
    - 40~200 mmHg (Pediatric)
    - 40~135 mmHg (Newborns)
  • NIBP DIA measurement range:
    - 10~210 mmHg (Adult)
    - 10~150 mmHg (Pediatric)
    - 10~95 mmHg (Newborns)
  • NIBP MAP measuring range:
    - 20~230 mmHg (Adult)
    - 20~165 mmHg (Pediatric)
    - 20~110 mmHg (Newborns)
  • NIBP measurement accuracy:
    - Average difference: ±5 mmHg
    - Standard deviation: 8mmHg
  • Measurement mode NIBP: Manual, Auto, STAT, Multicycle
  • Automeasurement intervals: 1-480 min
SpO2
  • Technique: Double wavelength optical method
  • Measuring range: 0%~100% ±2% for SpO2 range 70~100%
  • PR measuring range: 30~250bpm ±2bpm or ±2%, whichever is greater
  • Low Perfusion Efficiency: Up to 0.3%.
EtCO2 (optional, module and probes not included)
  • Technique: Infrared optical method
  • Sampling Mode: Sidestream or Mainstream
  • Measurement range: 0~150mmHg
  • Measurement Accuracy:
    - 0~40mmHg ±2mmHg
    - 41~70 mmHg ±5% of reading
    - 71~100 mmHg ±8% of reading
    - 101~150 mmHg ±10% of reading
  • Flow rate: 50ml/min ±10ml/min (Sidestream)
IBP extension
  • Technique: strain gauge transducer
  • Input sensitivity: 5 µV/V/mmHg
  • Measurement range: -50~300mmHg ±2% or ±4mmHg, whichever is greater
  • Measurement locations: ART, RAP, PA, LAP, CVP, ICP, AUXP1, AUXP2
  • Calibration: zero calibration
Other characteristics
  • Power Supply: AC 100-240V, 50/60Hz, 60VA
  • Built-in lithium battery: 11.1V/4400mAh
  • Screen: 15 inch TFT screen
  • Alarm method: 3-level audible-visual alarm
  • Network: Ethernet
  • Safety standards: IEC 60601-1

Accessories